Description
Intraoral X-Ray Sensors











Technical Specifications
|
Components
|
Specification
|
|
|---|---|---|
|
Sensor |
||
|
Sensor Technology |
CMOS |
|
|
Sensor matrix |
Size 1 : 1168 x 1562 pixels |
|
|
Sensor active |
Size 1 : 22.2 x 29.6 mm |
|
|
Gray scale |
4096 grey levels (12 bits) |
|
|
Dimension |
Size 1 : 27,6 mm(W) x 37,7 mm (H) x 7,3 mm (D) |
|
|
Weight |
Size 1 : 65 g |
|
|
Input voltage |
5V DC (USB interface) |
|
|
USB Interface |
USB 2.0 high Speed |
|
|
USB Connector |
Plug Type A |
Minimum Computer System Requirements
The computer and the peripheral equipment must conform to the IEC 60950 standard
|
Item
|
Item Display and Acquisition
|
|---|---|
|
CPU |
2 GHz Intel Duo Core |
|
RAM |
2 GB |
|
Hard disk drive |
1.2 GB for software installation |
|
Graphic board |
Nvidia/ATI based board with 256 MB of RAM |
|
Display |
1024 x 768 minimum screen resolution |
|
Operating system |
Windows 10 (64 bits) |
|
USB 2.0 |
2 ports available |
|
Ethernet |
100 BaseT |
|
CD/DVD drive |
DVD-ROM drive is required to install the product. |
|
Backup media |
Removable/portable, external hard disk drive |
Imaging Performance Information
Line Pair Resolution*: 5 lp/mm minimum.
Low Contrast Resolution*: all 4 low contrast elements are recognizable.
*Using a dental phantom for digital image acquisition that complies with the IEC61223-3-4:2000
standard and according to the minimum values requested in the DIN 6868-151.
Controlling the Image Quality
For optimum results, perform a control test of the image quality. To do this, see the User & Installation Guide.
Environmental Requirements
|
Ambient Operating Conditions
|
|
|---|---|
|
Temperatures |
10 – 40 °C |
|
RAM |
30 – 85% |
|
Atmospheric pressure |
700 – 1060 hpa |
|
Altitude |
Up to 3000 m |
|
Storage Conditions
|
|
|---|---|
|
Temperatures |
-10 – 60 °C |
|
RAM |
10 – 90% |
|
Atmospheric pressure |
700 – 1060 hpa |
|
Transport Conditions
|
|
|---|---|
|
Temperatures |
-10 – 60 °C |
|
RAM |
10 – 90% |
|
Atmospheric pressure |
700 – 1060 hpa |